Specifications

Electrical Shock Type BF Applied Part.
Environmental Conditions Transport and store 50°F to 131°F (10°C to 55°C). Operate 41°F to 104°F (5°C to 40°C). Humidity less than 90%. Atmospheric pressure from sea level to 9,842 feet (3,000m).
Ingress of Water Not protected against ingress of water.
Flammability Do not use around flammable gases, liquids or materials.
Mode of Operation Continuous.
Dimensions 6.1” x 3.5” x 1.9” (15.5 x 8.9 x 4.8cm)
Weight .4 lbs. (.18kg)
Cleaning Clean the case and lead clip wires as needed with an alcohol moistened cloth. Do not immerse in fluids.
Disposal Dispose of according to local, state and federal regulations. Remove battery before disposal.
Controls Two (dose and current).
Dose Range 0 to 80 mA-min.
Maximum Voltage 80V DC
Maximum Current 4.0 mA
Current Ramp Up Automatic (0 to 4.0 mA). Built-in option for manual override.
Current Ramp Down Automatic at end of treatment; paused or turned off by depressing knob. Built-in option for manual override.
Battery Use only 9V DC Alkaline. Ensure battery door is in place before starting treatment. Remove battery from unit when not in use.
Display Dose, Time Remaining, and Current (displayed simultaneously with interactive set-up).
Pause Feature YES (w/recalculation when restarted)
Visual Indicators Low battery and open circuit.
Audible Alerts Low battery, open circuit, and end-of treatment.
Auto Shut-Off Will automatically shut off after one (1) minute, if not in use.